|
Welcome
to the dissolved oxygen (D.O.) theory Page. Dissolved oxygen
from A to
Z.
Dissolved Oxygen:
The
air we breathe contains about 20% oxygen. Fish and other aquatic organisms
require oxygen as well. The term Dissolved Oxygen
(DO or D.O.) refers to the
amount of free oxygen dissolved in water which is readily available to
respiring aquatic organisms. State water quality standards often express
minimum concentrations of dissolved oxygen which must be maintained in
order to support life as well as be of beneficial use. Levels of dissolved
oxygen below 4-5 milligrams per liter affect fish health and levels below
2 milligrams per liter can be lethal to fish.
Additionally,
biochemical oxygen demand (BOD) is commonly used with reference to
effluent discharges and is a common, environmental procedure for
determining the extent to which oxygen within a sample can support
microbial life. The test for BOD is especially important in waste water
treatment, food manufacturing, and filtration facilities where the
concentration is crucial to the overall process and end products. High
concentrations of DO predict that oxygen uptake by microorganisms is low
along with the required break down of nutrient sources in the medium.
Basic
principles of Polagrography cell:
Liquid
and Air state of equilibrium is reached when the partial pressure of
oxygen, i.e. the part of the total pressure that is due to oxygen, is
equal in air and in liquid. The liquid is then saturated with oxygen.

Figure
1.1 Air and liquid oxygen equilibrium

Polargram:
When
an electrode of noble metal such as platinum or gold is made 0.6 to 0.8 V
negative with respect to a suitable reference electrode such as AgAgCl or
an calomel electrode in a neutral KCI solution (see Figure 1.2), the
oxygen dissolved in the liquid is reduce at the surface of the noble
metal.

Figure
1.2 Polarographhy diagram
This
above phenomenon can be observed from a current to voltage diagram called a
polarogram of the electrode. As shown in Figure 1.3a, the negative voltage
applied to the noble metal electrode (called the cathode) is increased,
the current increases initially but soon it becomes saturated. In this
plateau region of the polarogram, the reaction of oxygen at the cathode is
so fast that the rate of reaction is limited by the diffusion of oxygen to
the cathode surface. When the negative bias voltage is further increased,
the current output of the electrode increases rapidly due to other
reactions, mainly, the reduction of water to hydrogen. If a fixed voltage
in the plateau region (for example, - 0.6V) is applied to the cathode, the
current output of the electrode can be linearly calibrated to the dissolved oxygen
(Figure 1.3b). It has to be noted that the current is
proportional not to the actual concentration but to the activity or
equivalent partial pressure of dissolved oxygen, which is often referred
to as oxygen tension. A fixed voltage between -0.6 and -0.8 V is usually
selected as the polarization voltage when using Ag/AgCl as the reference
electrode or any other EID's dissolved oxygen electrodes.
Additionally
for physical and chemical correctness, partial pressure in a liquid
actually refers to the fugacity. In the pressure range relevant to the
measurements at hand, it is acceptable to equate the two values and this
allows us to restrict the following considerations to the partial
pressure. In dry, atmospheric air, the partial pressure of oxygen is
20.95% of the air pressure. This value is reduced over a water surface
because water vapor has its own vapor pressure and a corresponding partial
pressure.
Figure
1.3 (a) Current to voltage diagram at different oxygen tension; (b)
Calibration obtained at a fixed polarization voltage of –600 mV.
When
the cathode, the reference electrode, and the electrolyte are separated
from the measurement medium by a polymer membrane, which is permeable to
the dissolved gas but not to most of the ions and other species, and when
most of the mass transfer resistance is confined in the membrane, EID’s
electrode system can measure oxygen tension in various liquids. This is
the basic operating principle of the membrane covered polarographic
Dissolve oxygen probe (Figure 1.4).
The
basic principle underlying the electrochemical determination of oxygen
concentration is the use of membrane covered electrochemical sensors. The
main components of the sensors are the oxygen permeable membrane, the
working electrode, the electrolyte solution and a
possible reference electrode. A voltage is applied between the gold
(platinum or silver) cathode and the anode that consists of either lead or
silver (AgAgCl), and causes
the oxygen to react electrochemically. The higher the oxygen concentration
the higher the resulting electric current. The current in the sensor is
measured and, after calibration, converted into the concentration of dissolved oxygen.
If the anode is made of
silver, the meter applies the required voltage (polarographic sensor). If
it is made of lead, the sensor is self-polarizing, i.e. the voltage is
generated in the sensor by the electrodes themselves, comparable to the
process in a battery (galvanic sensor). The meter merely evaluates the
current.

Figure
1.4 Basic Polarographhy electrode

EID’s
polaragraphoc dissolved oxygen electrode picture:
EID’s
ELECTRODER - ABS
body Dissolved Oxygen Sensor (ADO)
EID’s
dissolved oxygen, Probe, polaragraphic, ABS body, 12mm * 120mm, with 10K
Negative Temperature Compensation
Figure
1.5 Basic Polarographhy-electrode

Electrode
reactions:
For
our polarographic electrodes, the reaction proceeds as follows:
H202
+ 2e- -> 20H-
The
reaction tends to produce alkalinity in the medium together with a small
amount of hydrogen peroxide.

Number
of electrons involved:
Two
principal pathways were proposed for the reduction of oxygen at the noble
metal surface. One is a 4-electron pathway where the oxygen in the bulk
diffuses to the surface of the cathode and is converted to H2O
via H2O2 (path a in Fig. 1.6). The
other is a 2-electron pathway where the intermediate H2O2 diffuses directly out of the
cathode surface into the bulk liquid (path b in Figure 1.6). The oxygen
reduction path may change depending on the surface condition of the noble
metal. This is probably the cause for time-dependent current drift of
polarographic sensors. Since the hydroxyl ions are constantly being
substituted for chloride ions as the reaction starts, KCI or NaCl has to
be used as the electrolyte. When the electrolyte is depleted of Cl-,
it has to be replenished.
2e-
2e- à (a) à H2O2
O2 à
O2 à
H2O2
Diffusion
à (b) à H2O2
Figure
1.6 alternative pathway of oxygen reduction at cathode surface

Calibration:
Calibration
must be carried out for dissolved oxygen measurements on a regular base.
This is because the measuring process consumes the electrolyte solution in
the sensor head, as shown by the electrode reactions presented above. The
ions of the electrolyte solution bind the released metal ions, thereby
changing the composition of the solution. The recommended calibration
period depends on the oxygen sensor used and ranges from one week for
pocket instruments to 1-2 months.
Each
linear calibration function is defined by at least two points. For dissolved oxygen
measurements with EID meter and/or logger, one of the
points on the line is the zero point of the sensor. At the zero point, the
sensor signal obtained in the absence of oxygen lies below the resolution
of the sensor. This point is called the zero-current point of the sensor.
The second point of the calibration line can be set as required. Its
position is based on the fact that, in a state of equilibrium, the partial
pressure of oxygen in liquid and air is equal.
Figure
1.4b Two-point calibration
The
rate at which oxygen enters a dissolved oxygen probe is a function of:
As
described above calibration routines for dissolved oxygen probes use a two
point linear calibration where one point is at zero mg/L oxygen and the
second point is at saturation or equilibrium with the atmosphere, C*
. The
zero measurement is not zero volts due to the conductivity of the
electrolyte between the electrodes as well as any errors in the analog
signal conditioning circuit. For the circuit and probe system used in the
Environmental lab the zero measurement
is approximately 1 mV (where approximately 200 mV corresponds to
saturation levels of oxygen) and hence the zero measurement is not
significant. Thus a single point calibration is used.
C*
is a function of the atmospheric pressure and temperature. The functional
relationship with temperature is implemented using a lookup table (based
on equilibrium at atmospheric pressure) with interpolation. The effect of
atmospheric pressure is implemented as shown (Equation 1 below).
P
C*
=------- f (T)
Patm
The
permeability of the membrane increases about 5% per C degree. I chose to
use 25 C as the reference temperature and thus Kmembrane(Tref)
has a value of 1. The following (equation 2) creates a coefficient
that describes this variation.
Kmembrane(T)=Kmembrane(Tref)e0.05(T-Tref)
The
slope of the linear fit (k) can be calculated after the voltage
corresponding to saturation oxygen is measured (Equation 3 below).
C*cal Kmembrane
C
=------------------
V*cal
The
slope coefficient is placed in the polynomial array.
The
equation for the dissolved oxygen concentration illustrates that the
predicted concentration is a function of sample temperature because Kmembrane
varies with temperature. The coefficient, k, should be independent
of temperature but will vary as a membrane fouls (Equation
4 below)
KV
C=-----------
Kmembrane
Pressure
The
constituents of air have been well defined, and it is known that air
contains 20.946% oxygen. Since the total pressure in the air is the sum of
all of the partial pressures (Dalton’s Law), an atmospheric pressure of
760 millimeters Mercury (mmHg) in dry air will contain a partial pressure
of oxygen (pO2) of approximately 159 mmHg (760 mmHg * 0.20946).
Changes in atmospheric pressure will cause a directly proportional
change in the partial pressure of oxygen in the air.
Atmospheric pressures will vary depending upon altitude and local
weather conditions. Some
average pressures for varying altitudes are listed in Table 1 bellow.
The
relationship between oxygen partial pressure and total atmospheric
pressure should be understood and incorporated into the air calibration in
order to minimize calibration error, which could be as high as 5-10%
dependent upon altitude and local weather conditions.
Most dissolved oxygen meters that have any sort of advanced air
calibration (such as temperature compensation, which will be discussed in
a later section) will be based upon an atmospheric pressure of 760 mmHg.
Most tables of oxygen solubility are referenced to this value.
Because of the change in oxygen partial pressure with changes in
atmospheric pressure, a correction must be made when the pressure varies
from this value. A simple
means of incorporating pressure changes is listed in the “correction
factor” shown in Table 1 bellow. The
value listed is a rough multiplier, which can be used once the initial
oxygen concentration is determined based upon temperature and relative
humidity. A more accurate
calculation for incorporating pressure will be discussed after relative
humidity and temperature effects are investigated.
Some
EID’s dissolved oxygen meters contain a pressure sensing device which
provides compensation for pressure effects when an air calibration is
performed. If you use our
electrode on not-EID-meter, since most meters do not have this, it is
usually necessary to note the average pressure in the local vicinity of
the probe, which will be mostly altitude-based, and adjust the calibration
using the simple correction factor or the more complex calculation
performed later. A mercury
barometer located in the immediate vicinity of the meter will give a
relatively accurate measurement of the local atmospheric pressure if an
older meter with no pressure sensor is used.
|
Altitude
(ft)
|
Pressure
(mm
Hg)
|
Correction
Calibration
Correction
Factor
|
|
-540
|
775
|
1.02
|
|
Sea
Level
|
760
|
1
|
|
500
|
746
|
0.98
|
|
1000
|
732
|
0.96
|
|
1500
|
720
|
0.95
|
|
2000
|
707
|
0.93
|
|
2500
|
694
|
0.91
|
|
3000
|
681
|
0.9
|
|
3500
|
668
|
0.88
|
|
4000
|
656
|
0.86
|
|
4500
|
644
|
0.85
|
|
5000
|
632
|
0.83
|
|
5500
|
621
|
0.82
|
|
6000
|
609
|
0.8
|
Table
1: Oxygen Value Corrected for Pressure (25 °C)
Relative
Humidity and temperature effect and Temperature compensation
If
desired the Eid's probes can be coupled with a temperature thermistor (10K
Ohms) to achieve temperature compensation since Kmembrane
varies with temperature.
The
discussion of pressure effects were based upon atmospheric pressure with
dry air (no moisture content). Whenever
air contains a certain amount of moisture, the atmospheric pressure
contains another source of partial pressure -- water vapor.
If a comparison of the oxygen partial pressure in air with 100%
relative humidity and air with 0% relative humidity is done while both are
at the same atmospheric pressure, the air with 100% relative humidity will
have a lower oxygen partial pressure due to the presence of the water
vapor pressure (pH2O). Water
vapor pressure in air varies with temperature, and is well defined.
The effect of temperature on oxygen partial pressure in moist air
is such that higher temperatures yield lower oxygen partial pressure,
while lower temperatures yield higher pressures.
Note that the effects of relative humidity and temperature can
cause errors when air calibration is performed in dry air, since most of
the current tables and meter temperature compensations are based on air
containing 100% relative humidity. Table
2 bellow shows both the oxygen concentration, which is linear with the
partial pressure of oxygen, that
would be present at 100% relative humidity and 0% relative humidity.
The values only differ by a few percent in ambient air conditions,
and thus is generally ignored. Most dissolved oxygen meters have
temperature compensation for air at 100% relative humidity, and no manual
correction is necessary. However,
many older meters do not have temperature compensation included, and
therefore this calculation must be done manually.
If temperature is not compensated for in the calibration, the error
can be as much as 20 to 30 % for every 10 degrees difference from 25 °C,
and therefore temperature compensation is standard on most dissolved
oxygen meters today. Since
the effects of relative humidity is minimal at all but the highest
temperatures, no current dissolved oxygen meters incorporate any kind of
relative humidity sensing device.
In
order to ensure an accurate temperature and current reading, the probe
must be exposed to the air for enough time to allow thermal equilibrium to
occur. There are often
significant temperature differences between the process water and the
ambient air. Larger
temperature gradients between the two necessitate additional time for
thermal equilibrium to take place. For
instance, a 20 °C
difference between ambient air and process water can cause a calibration
delay of about 30 minutes in many probes for the probe to fully
equilibrate to ambient temperature. Since
most temperature gradients will not be this large, allowing approximately
15 minutes is usually a safe assumption.
It is common for users to calibrate the unit before the dissolved
oxygen meter is reading the stabilized temperature and current value,
which can cause significant error since a difference of even 5 °C
from actual can cause the reading be off by 5 to 10%.
It is often useful to have a calibrated temperature sensor,
accurate to 1 °C
or better, at the calibration location to know when the probe temperature
is reading the correct ambient air temperature.
It
is useful to have an equation which can be used to determine oxygen
concentrations in air based upon temperature, relative humidity, and
pressure. Since the full equation is quite lengthy and complex, two easier
versions are presented to the user, along with Table 2 bellow, to
determine the correct oxygen concentration in air.
Equation 5 bellow should be used with air with 100% relative
humidity, and Equation 6 should be used for air with 0% relative humidity.
Equation
5 (100% Relative Humidity): OS =
(OS’) * (P - p) / (760 - p)
where:
OS
= Oxygen solubility at barometric pressure of interest
OS’
= Oxygen in saturation at one atmosphere (760 mmHg) at a given temperature
P
= Barometric pressure of interest
p
= Vapor pressure of water at the temperature of interest
Example
1:
The
user wishes to calibrate a dissolved oxygen probe in air at an altitude of
3500 feet. The temperature is
30 °C,
and the relative humidity is 100%.
At
an altitude of 3500 feet, the atmosphere pressure will usually be about
668 mmHg (Table 1 above). The
sample temperature is 30 °C,
and the relative humidity is 100%. From
water vapor pressure tables, the water vapor pressure at 30 °C
is 31.8 mmHg. The oxygen saturation
level at 760 mmHg and 30 °C
is 7.54 ppm (Table 2 bellow). Substituting
these values in the above (equation 5) gives the following:
OS
= (7.54)
* (668 - 31.8) / (760 - 31.8) =
6.59 ppm
Example
2:
Assume
the same conditions as in example 1, but with a relative humidity of 0%.
In this case, the value used for the oxygen saturation level would
be 7.87 (Table 2 bellow), not 7.54. The
calculation will change since there will be no water vapor pressure.
Equation
6 (0% Relative Humidity): OS =
(OS’) * (P) / (760 mmHg)
Substituting
the above values into the equation yields the following:
OS
= 7.87 * (668) / (760) = 6.92 ppm
Note:
that the multiplier of (668) / (760) is actually the simplified correction
factor listed in Table 1 above for an altitude of 3500 feet (0.88). Table
3 bellow lists calibration values for varying temperatures pressures at
relative humidity levels of 100%.
|
Temperature
(Celsius)
|
DO (100% R.H.)
(ppm, mg/L)
|
DO (0% R.H.)
(ppm, mg/L)
|
|
0
|
14.6
|
14.66
|
|
1
|
14.19
|
14.26
|
|
2
|
13.81
|
13.89
|
|
3
|
13.44
|
13.53
|
|
4
|
13.09
|
13.18
|
|
5
|
12.75
|
12.85
|
|
6
|
12.43
|
12.54
|
|
7
|
12.12
|
12.23
|
|
8
|
11.83
|
11.94
|
|
9
|
11.55
|
11.66
|
| |